| Model Number |
138105 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During incoming inspection, the distributor rejected this device, catalog # 138105, lot 202004271, for a possible breach of sterility due to an insufficient heatseal.There was no contact with any patient as this was found during incoming inspection in (b)(6).The completion of the device evaluation confirmed an insufficient heatseal; therefore, this complaint meets the criteria for a reportable event.Due to the breach in sterility, this report is being raised as a malfunction with potential for injury upon reoccurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation of the customer's complaint of an insufficient heat seal leading to a breach in sterility has confirmed the reported issue.Conmed received two 138105 returned unopened in original packaging.The lot numbers of the devices were verified via returned packaging.A visual inspection was performed, one devices package was not sealed at one end and the second device was encroaching into the seal.A functional inspection and dye leak testing were performed which indicated that the packaging did not have an insufficient heat seal on the package with the device encroaching the seal.The manufacturing documents from the device history record have been reviewed and found no abnormalities that would contribute to this issue.A two-year lot history review shows this is the only complaint for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 143 complaints, regarding 713 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.0004.Conmed encourages the inspection and/or test of all medical equipment prior to use to ensure all devices are functioning as expected.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Search Alerts/Recalls
|