SMITH & NEPHEW, INC. JOURN ART INS BCS STD 5-6 LT13; PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMICONSTRAINED,CEMENTED,POLYMERMETALPOLYMER Back to Search Results
Model Number 74023264
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, after a tka surgery had been performed, a post of the journ art ins bcs std 5-6 lt13 broke.This adverse event was solved by a revision surgery performed on (b)(6) 2021, in which the broken device was replaced.The current health status of patient is unknown.
 
Additional Manufacturer Narrative
Internal reference number: (b)(4).
 
Event or Problem Description
It was reported that, after a tka surgery had been performed, a post of the journ art ins bcs std 5-6 lt13 broke.This adverse event was solved by a revision surgery performed on (b)(6) 2021, in which the broken device was replaced.The patient has been discharged from hospital and is recovering at home.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, this case reports, a revision was performed after a tka surgery (date unspecified), to retrieve and replace a tibial insert where the post was found to be broken.Reportedly the patient¿s current health status is ¿the patient has been discharged from hospital and is recovering at home.¿ no additional requested clinically relevant information was provided.Without the requested information, the root cause of the reported ¿post of jrny insert broke off,¿ cannot be definitively concluded.The patient impact beyond the device breakage and revision cannot be determined, and the patient¿s current health status is ¿discharged from hospital and is recovering at home.¿ therefore, no further clinical medical assessment can be rendered.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device, but no similar events for the batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.The instructions for use documents revealed this failure mode was previously identified.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to traumatic injury or surgical technique.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURN ART INS BCS STD 5-6 LT13
Common Device Name
PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMICONSTRAINED,CEMENTED,POLYMERMETALPOLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13205833
Report Number1020279-2022-00142
Device Sequence Number11158118
Product Code JWH
UDI-Device Identifier03596010555236
UDI-Public03596010555236
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K042515
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year1900
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2017
Device Model Number74023264
Device Catalogue Number74023264
Device Lot Number07BM03453
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/23/2021
Supplement Date Received by Manufacturer01/17/2022
02/10/2022
Initial Report FDA Received Date01/10/2022
Supplement Report FDA Received Date01/19/2022
02/11/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/04/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-