Initial reporter occupation: non-healthcare professional.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
|
Investigation: investigation is reopened due to additional information provided.The following allegations have been investigated: vena cava perforation, tilt, fear, stress, anxiety, loss of enjoyment of life, mental issues, emotional issues.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported fear, stress, anxiety, loss of enjoyment of life, mental issues, emotional issues is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog and lot number are unknown.The alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
Patient allegedly received an implant on (b)(6) 2010 via the right common femoral vein due to deep vein thrombosis (dvt) and gastrointestinal (gi) bleeding.Patient is alleging vena cava perforation, and tilt.Patient also alleges "severe fear, stress, anxiety and loss of enjoyment of life", mental issues and emotional issues per computed tomography (ct) report, "an ivc filter is at the l1-2 level." per ct report, "the filter apex begins at the level 1.6 cm below left renal vein entrance.The apex of the filter protrudes outside the medial wall of the cava." "the filter is tilted so that the apex contacts the medial wall of the cava.The long axis is parallel to the cava.The filter shows perforation of the inferior struts into peri caval fat all the way around but these struts do not penetrate into any other structure.".
|