| Model Number |
M00539200 |
| Medical Device Problem Codes |
Break (1069); Positioning Failure (1158); Entrapment of Device (1212); Activation, Positioning or Separation Problem (2906); Material Deformation (2976); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter phone: (b)(6).(b)(4).The returned flexima biliary was analyzed, and a visual evaluation noted that microscopic inspection of the suture hole of the push catheter was torn.It was observed that the guide catheter was kinked and it was detached, the push catheter presented some kinks as well.The guide catheter had one guidewire inside of it, during functional inspection it could be removed without issues.The stent and suture was not returned.The reported event was confirmed.According to product analysis, the push catheter kinked may be related to user manipulation during the preparation of the device, some technique applied during the procedure and/or even tortuous anatomy could have contributed on this issue, once the push catheter was kinked the user experienced difficulties to release the stent and this may have taken the user to apply a different technique and/or apply an extra force, as consequence the guide catheter got kinked and finally detached, also during the attempts of deployment the suture hole got torn due these conditions.Therefore, adverse event related to procedure is selected as the most probable root cause for the complaint.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a flexima biliary stent was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed in the common bile duct on (b)(6) 2021 for the treatment of secondary sclerotic cholangitis.During the procedure and inside the patient, the flexima stent failed to deploy.The guidewire was completely removed for deployment, but the suture did not peel away, preventing stent deployment.A second flexima biliary stent was attempted to be used, but the same issue occurred.A third flexima biliary stent device was successfully used to complete the procedure.There was no serious injury, nor were there any adverse patient effects reported as a result of this event.This event has been deemed a reportable event based on the investigation finding of guide catheter broken.
|
| |
|
Search Alerts/Recalls
|
|
|