| Model Number |
367297 |
| Medical Device Problem Code |
Difficult or Delayed Activation (2577)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Oem manufacture: the manufacturing location for this product is (b)(4).This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.Investigation summary: bd had not received samples or photos from the customer for investigation.Therefore, 50 retention samples from bd inventory were evaluated by visual examination and functional testing and no issues were observed relating to dull/does not easily retract as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality non-conformances during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure modes.Bd was not able to identify a root cause for the indicated failure modes.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that bd vacutainer® safety-lok¿ blood collection set had a difficult to activate safety feature.The following information was provided by the initial reporter: "it was reported that the needle is dull and it does not retract easily.".
|
| |
|
Additional Manufacturer Narrative
|
|
Oem manufacture: the manufacturing location for this product is (b)(4).This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.Investigation summary: bd had not received samples or photos from the customer for investigation.Therefore, 50 retention samples from bd inventory were evaluated by visual examination and functional testing and no issues were observed relating to dull/does not easily retract as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality non-conformances during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure modes.Bd was not able to identify a root cause for the indicated failure modes.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that bd vacutainer® safety-lok¿ blood collection set had a difficult to activate safety feature.The following information was provided by the initial reporter: "it was reported that the needle is dull and it does not retract easily.".
|
| |
|