| Model Number |
IPN040153 |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Event or Problem Description
|
|
Epidural catheter was inserted successfully by anesthetist on call.The anesthetist then placed the syringe adapter on the end of the catheter and proceeded to instill the medication but was unable to push any medication into the epidural catheter.The doctor removed the catheter and was still unable to push any medication into it.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).A device history record review was performed on the epidural catheter with no relevant findings.A corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed on the epidural catheter with no evidence to indicate a manufacturing related issue.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without a sample.No further action is required at this time.
|
| |
|
Event or Problem Description
|
|
Epidural catheter was inserted successfully by anesthetist on call.The anesthetist then placed the syringe adapter on the end of the catheter and proceeded to instill the medication but was unable to push any medication into the epidural catheter.The doctor removed the catheter and was still unable to push any medication into it.
|
| |
|