(b)(4).Date sent: 05/09/2022.D4: batch # 324a73.Device analysis: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the tx30v device arrived with no apparent damaged and with a reload loaded on the device.In addition, a tyvek was received along with the device.The reload was received fully fired.The device was tested for functionality with a test reload and it fired and formed all the staples as intended.The staple line was complete, and the staples meet the staple form release criteria.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: prior to opening the instrument and releasing the tissue, the edge of the reload or the side of the anvil may be used as a guide to transect tissue (not vessels) or to excise tissue which is protruding through the jaws.This aids in cutting at a proper distance from the staple line.Open the jaws by pushing the release button and releasing the grasp of the closing trigger.The triggers and jaws will fully open, releasing the tissue.Remove the instrument.The event could not be confirmed as the device performed as expected and during the functional testing, the device opened and closed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|