ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL VASCULAR LIN STAP; STAPLE, IMPLANTABLE Back to Search Results
Model Number TX30V
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/21/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Did the device open? was the device difficult to close on the tissue? was the device difficult to fire? was the device difficult to open? on which firing did this occur? did the tissue retaining pin lock into the correct position?.
 
Event or Problem Description
It was reported that during an unknown procedure, surgeon could not open device after firing complete.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 05/09/2022.D4: batch # 324a73.Device analysis: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the tx30v device arrived with no apparent damaged and with a reload loaded on the device.In addition, a tyvek was received along with the device.The reload was received fully fired.The device was tested for functionality with a test reload and it fired and formed all the staples as intended.The staple line was complete, and the staples meet the staple form release criteria.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: prior to opening the instrument and releasing the tissue, the edge of the reload or the side of the anvil may be used as a guide to transect tissue (not vessels) or to excise tissue which is protruding through the jaws.This aids in cutting at a proper distance from the staple line.Open the jaws by pushing the release button and releasing the grasp of the closing trigger.The triggers and jaws will fully open, releasing the tissue.Remove the instrument.The event could not be confirmed as the device performed as expected and during the functional testing, the device opened and closed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 
6107428552
MDR Report Key13206862
Report Number3005075853-2022-00185
Device Sequence Number4564433
Product Code GDW
UDI-Device Identifier10705036003335
UDI-Public10705036003335
Combination Product (Y/N)N
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Model NumberTX30V
Device Catalogue NumberTX30V
Device Lot Number196A03
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/22/2021
Supplement Date Received by Manufacturer04/13/2022
Initial Report FDA Received Date01/10/2022
Supplement Report FDA Received Date05/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-