• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. SELF-STARTING, VARIABLE SCREW; SIZE Ø4.0X14 MM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. SELF-STARTING, VARIABLE SCREW; SIZE Ø4.0X14 MM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY Back to Search Results
Model Number 8801-04014DA
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device remains implanted in patient.
 
Event or Problem Description
It was reported that two ozark screws at c6 fractured post-operatively.Revision surgery has not been scheduled at this time.This report will capture the first of two screws.
 
Event or Problem Description
It was reported that two ozark screws at c6 fractured post-operatively.Revision surgery has not been scheduled at this time.This report will capture the first of two screws.
 
Additional Manufacturer Narrative
Visual, dimensional, material and functional analysis could not be performed as the screw associated with this record was not returned.Device history records could not be reviewed as a valid lot number was not provided and could not be obtained.Complaint history records were reviewed for this catalog number, no similar complaints were identified.Initial report states that two ozark screws at c6 fractured post-operatively.No x-ray or product return was received.No complications during initial procedure and no revision surgery has been performed.Drill was used to prep screws holes, no tap was used.No reported patient information or post-op activity.Ozark view and guide surgical technique was reviewed: ¿these internal fixation devices are load sharing devices which maintain alignment until healing occurs.If healing is delayed or does not occur, the implant could eventually break, bend, or loosen.Physical activity and load bearing have been implicated in premature loosening, bending or fracture of internal fixation devices in the absence of complete bone healing.¿ ¿adequately instruct the patient.Postoperative care and the patient¿s ability and willingness to follow instructions are two of the most important aspects of successful healing.¿ due to lack of product information, device not returned and with no x-ray provided an exact cause of the reported event of the screw fracture cannot be conclusively determined at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF-STARTING, VARIABLE SCREW; SIZE Ø4.0X14 MM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key13207074
Report Number3004774118-2022-00015
Device Sequence Number1686859
Product Code KWQ
UDI-Device Identifier10888857393363
UDI-Public10888857393363
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K172104
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8801-04014DA
Device Catalogue Number8801-04014DA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/14/2021
Supplement Date Received by Manufacturer03/22/2022
Initial Report FDA Received Date01/10/2022
Supplement Report FDA Received Date04/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-