| Model Number |
AX-80075 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zyno medical is waiting for the arrival of the affected device.
|
| |
|
Event or Problem Description
|
|
On 12/23/2021, zyno medical received a complaint from a user facility representative.It was reported that "an administration set model ax-80075 lot 2109002 set had a particle in the drip chamber after completing a chemo therapy treatment." a patient was involved but was not harmed or injured.The device operator was a registered nurse.The medication being infused was unknown.The contract manufacturer of the affected device is podo xingda (b)(4) medical co.Ltd.
|
| |
|
Additional Manufacturer Narrative
|
|
The service provider provided the investigation report on 03/04/2022.The actual device was inspected.A small piece of circle shaped white object was observed inside the drip chamber.The reported issue is confirmed but the root cause cannot be identified.
|
| |
|
Event or Problem Description
|
|
This is a follow-up for the initially filed mdr (3006575795-2021-00059).
|
| |
|
Event or Problem Description
|
|
This is the second follow-up for the initially filed mdr (3006575795-2021-00059).
|
| |
|
Additional Manufacturer Narrative
|
|
The contract manufacturer provided the investigation report on 05/19/2022.The actual device was inspected.The white particle was confirmed to be an air intake film, which is a component of the administration set.It is believed to be a manual assembly problem during the manufacturing process.Operators at the production site were notified about this complaint and they have been required to pay attention during the production operation.
|
| |
|
Search Alerts/Recalls
|