• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. NS VIS ADPT GUIDE JII KIT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. NS VIS ADPT GUIDE JII KIT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number V0200112
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Ecchymosis (1818); Pain (1994); Deformity/ Disfigurement (2360); Ambulation or Postural Difficulties (2544); Swelling/ Edema (4577); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that, after a tka surgery, the knee of the patient ended up in 10 degrees of valgus.According to surgeon, both blocks from ns visionaire adaptive guide kit journey ii fit well during the procedure.The tibia alignment rod was used to place block.The cuts were made.Once the implants were cemented, the surgeon used alignment rods to measure and noticed that the knee was in valgus.The patient current status is unknown.
 
Additional Manufacturer Narrative
Additional information: h6 (component code, type of investigation, investigation findings, investigation conclusions).Results of investigation: given the nature of the alleged incident, the device, could not be returned for evaluation.This issue was evaluated through our internal quality process and determined the reported event could not be confirmed.The clinical/medical investigation concluded that, the reported ¿knee of the patient ended up in 10 degrees of valgus,¿ cannot be confirmed based on the analysis of the x-rays and the single intraop image provided.The engineering evaluation revealed findings were inside acceptance criteria.Per the office note, the 01-04-22 x-rays show ¿a right tk prosthesis well aligned, intact with no evidence of loosening¿ or tissue/bony abnormalities and was not identified in the operative report.It was communicated that although the surgeon was not satisfied with the outcome ¿initially¿ in the or, the surgeon was satisfied ¿after seeing him in clinic¿.The patient impact beyond the reported ¿patient ended up in 10 degrees of valgus¿, post-pain, effusion, and limp cannot be determined.It is unknown if any future interventions are planned or scheduled.No further medical assessment can be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the previous 24 months did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to surgical technique used or user/procedural variance.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Corrected data: h6 (health effect - clinical code, health effect - impact code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NS VIS ADPT GUIDE JII KIT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13208756
Report Number1020279-2022-00147
Device Sequence Number14619371
Product Code JWH
UDI-Device Identifier00885556658406
UDI-Public00885556658406
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K170282
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberV0200112
Device Catalogue NumberV0200112
Device Lot Number00220339
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2021
Supplement Date Received by Manufacturer09/30/2022
Initial Report FDA Received Date01/10/2022
Supplement Report FDA Received Date10/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-