Model Number D7TDL252RT
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Note: the section ¿initial reporter¿ information is currently unavailable.The only information available is the country of event.Follow up is in progress to obtain this information.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) female patient underwent an atrioventricular nodal reentry tachycardia (avnrt) ablation procedure with a celsius¿ electrophysiology catheter and the catheter tip broke.It was reported that there was a fracture between the electrode tip.The surgery was not delayed due to the reported event.They changed the catheter when the event occurred and the procedure was successfully completed.Fragments were not generated.No other medical intervention was required.There was no patient consequence.Multiple attempts have been made to obtain clarification to this complaint.However, no further information has been made available.
 
Additional Manufacturer Narrative
Note: the section ¿initial reporter¿ information is currently unavailable.The only information available is the country of event.Follow up is in progress to obtain this information.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) female patient underwent an atrioventricular nodal reentry tachycardia (avnrt) ablation procedure with a celsius¿ electrophysiology catheter and the catheter tip broke.It was reported that there was a fracture between the electrode tip.The surgery was not delayed due to the reported event.They changed the catheter when the event occurred and the procedure was successfully completed.Fragments were not generated.No other medical intervention was required.There was no patient consequence.Multiple attempts have been made to obtain clarification to this complaint.However, no further information has been made available.
 
Additional Manufacturer Narrative
On (b)(6)2022, additional information was received indicating the damage resulted in wires being exposed but no lifted or sharp rings.There was no resistance or difficulty during insertion or removal of the catheter since the device was not used on the patient due to the damage being found when the package was open.It has not yet been insert into the patient.The lot number for the device was reported to be 30465035m.Device evaluation details: on (b)(6)2022, the product was returned to biosense webster inc.(bwi) for evaluation and the evaluation has been completed.Bwi conducted visual analysis of the returned device.The visual analysis of the returned sample revealed that the tip was found slightly bent.The ¿fracture¿ condition is actually part of the catheter¿s design and called the anchor window.A tilt test was performed, in accordance with bwi procedures.The catheter was found in the specification.In addition, the customer provided a picture of the complaint device to aid in the investigation.The picture shows the catheter tip, but does not provide sufficient information related to the reported event and therefore no result can be obtained from it.As part of bwi¿s quality process, all devices are manufactured, inspected, and released to approved specifications.The events described could not be confirmed and no issues were observed.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.The instructions for use (ifu) contain the following warnings and precautions: always pull the thumb knob back to straighten the catheter tip before insertion or withdrawal of the catheter.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Additional Manufacturer Narrative
On 10-jan-2022, the details related to the initial reporter were received.As such, the appropriate fields in section e have been populated: (b)(6).Thailand please note, a character limit in field e1.Initial reporter address line 1 prompted the information to be minimized as such: 106 moo 4 lardkrabang ind esta if additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELSIUS¿ ELECTROPHYSIOLOGY CATHETER
Common Device Name
ELECTRODE, PERCUTANEOUS, CONDUCTION TISSUE ABLATION
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
gabriel alfageme
31 technology dr
irvine, CA 92618
9497898687
MDR Report Key13211610
Report Number2029046-2022-00064
Device Sequence Number13874713
Product Code LPB
UDI-Device Identifier10846835000412
UDI-Public10846835000412
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
P950005/S1
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/02/2023
Device Model NumberD7TDL252RT
Device Catalogue NumberD7TDL252RT
Device Lot Number30465035M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2021
Supplement Date Received by Manufacturer01/10/2021
01/12/2021
Initial Report FDA Received Date01/10/2022
Supplement Report FDA Received Date01/12/2022
02/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DEF 7F,T,CC,4P,D,LE,252,RL,115
Patient Age70 YR
Patient SexFemale
-
-