• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number SMXP150RH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 11/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: material rupture.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) year old caucasian female patient, who's undergone primary breast reconstruction with two 750cc artoura breast tissue expanders, suffered right breast implant deflation, post-procedure.As a result, the patient underwent removal and replacement with an unspecified silicone implant on (b)(6) 2021.
 
Additional Manufacturer Narrative
On (b)(6) 2022, the mentor failure analysis lab received the device for evaluation.On (b)(6) 2022, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection and leak testing of the returned tissue expander.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the artoura ss, t exp, hp,750cc tissue expander.Leak testing was performed, in accordance with mentor procedures, and a tear was noted on the anterior aspect measuring less than 0.1 cm.As the authorization form for examination was not received the product evaluation lab couldn´t identify a conclusion due to the specific nature of this deflation.The evaluation was limited to non-destructive testing.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of tissue expander include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTOURA BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
gabriel alfageme
3041 skyway circle north
irving, TX 75038
9497898687
MDR Report Key13212452
Report Number1645337-2022-00302
Device Sequence Number16520401
Product Code LCJ
UDI-Device Identifier00081317024347
UDI-Public00081317024347
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/17/2023
Device Model NumberSMXP150RH
Device Catalogue NumberSMXP150RH
Device Lot Number7697751
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/14/2021
Supplement Date Received by Manufacturer01/25/2022
Initial Report FDA Received Date01/10/2022
Supplement Report FDA Received Date02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-