MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LITE; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LITE; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 70805-70
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 12/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At this time, product has not yet been returned.An extended investigation has been performed for the reported complaint, and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history record) for the freestyle lite meter and freestyle strips were reviewed, and the dhrs showed the freestyle lite meter and freestyle strips passed all tests prior to release.Retain testing was not performed for freestyle strips as the test strips expired on 31 aug 2020.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The date the incident occurred is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
A customer reported a delivery delay with a replacement adc blood glucose meter.The replacement device was issued as customer had reported low glucose results.Due to delivery delay, customer reported being hospitalized due to high blood glucose as she had no method to monitor glucose.The customer did not provide treatment information but did state she was prescribed insulin after hospitalization.There was no report of death or permanent injury associated with this event.
 
Additional Manufacturer Narrative
The reported meter has been returned and investigated.Visual inspection has been performed on the meter no issues were observed.Meter powered on with button and retained test strips.Control solution testing has been performed and all results were within range specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
A customer reported a delivery delay with a replacement adc blood glucose meter.The replacement device was issued as customer had reported low glucose results.Due to delivery delay, customer reported being hospitalized due to high blood glucose as she had no method to monitor glucose.The customer did not provide treatment information but did state she was prescribed insulin after hospitalization.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LITE
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key13213604
Report Number2954323-2022-01045
Device Sequence Number4745109
Product Code NBW
UDI-Device Identifier00699073708274
UDI-Public00699073708274
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K092602
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date08/31/2020
Device Model Number70805-70
Device Catalogue Number70827
Device Lot Number1061979
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2021
Supplement Date Received by Manufacturer02/21/2022
Initial Report FDA Received Date01/10/2022
Supplement Report FDA Received Date03/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/12/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age84 YR
Patient SexFemale
Patient Weight75 KG
Patient EthnicityNon Hispanic
Patient RaceAmerican Indian Or Alaskan Native
-
-