• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY Back to Search Results
Model Number 295025-001
Medical Device Problem Codes Battery Problem (2885); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The freedom onboard battery will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.Ce 5735 (1 fo 2) initial.
 
Event or Problem Description
The reported issues involve the following syncardia temporary total artificial heart (tah-t) system components and are reported under two separate medical device reports: (1) freedom onboard battery s/n (b)(4) (mfr report # 3003761017-2022-00001 and (2) freedom onboard battery s/n (b)(4) (mfr report # 3003761017-2022-00002).The customer, a syncardia certified hospital, reported that the patient was at home when the freedom driver started overheat and exhibited an alarm.The patient also reported that he exchanged the freedom driver onboard batteries and the driver stopped alarming.The patient put the freedom onboard batteries, that were in the driver, into the battery charger and reported they would not fully charge.There was no reported adverse patient impact.
 
Additional Manufacturer Narrative
The freedom onboard battery was evaluated per the freedom onboard battery evaluation procedure and passed all sections.During investigation testing, there was no evidence of a device malfunction and the battery performed as intended.The freedom driver used at the time of the reported issue was not returned and therefore could not be included in this investigation.The root cause of the customer-reported freedom driver alarm when the freedom onboard battery was inserted could not be conclusively determined.This issue will be tracked and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1 (1 of 2).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM ONBOARD BATTERY
Common Device Name
BATTERY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
sagar pimpalwar
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key13214757
Report Number3003761017-2022-00001
Device Sequence Number12095612
Product Code LOZ
UDI-Device Identifier00858000003084
UDI-Public(01)00858000003084
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P030011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number295025-001
Device Catalogue Number295025-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/29/2021
Supplement Date Received by Manufacturer12/29/2021
Initial Report FDA Received Date01/10/2022
Supplement Report FDA Received Date08/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/03/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age47 YR
Patient SexMale
-
-