ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA22-70/I20-30
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 09/25/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was initially implanted with an afx2 bifurcated stent graft and a suprarenal stent graft extension to treat an abdominal aortic aneurysm (aaa) on (b)(6) 2016.Approximately four (4) years post initial procedure, the patient presented with a type ii and possible ib endoleaks.The physician elected to treat the patient by relining with an afx2 bifurcated device, an afx suprarenal, and two (2) ovation ix extenders on (b)(6) 2020.Reference mfr.Report # 2031527-2020-00106 for this previously reported event.After approximately two (2) years post-secondary procedure, the patient presented with an aneurysm enlargement.The patient is currently being monitored.
 
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.If additional information is obtained at a later time that is pertinent to this event, a follow-up report will then be submitted.Device iteration is afx2.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident-related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the sac growth of 6.6mm was confirmed.This is consistent with the reported adverse event/incident.Procedure-related harms, device, user, procedure, or anatomy-relatedness of this complaint could not be determined with the medical records available for review.It is unclear what the source of the sac growth was.The final patient status was reported to be under surveillance.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.Corrections: g3: date of received by manufacturer - updated.H6: investigation finding codes: remove code 3233.H6: investigation conclusion codes: remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key13216201
Report Number2031527-2022-00003
Device Sequence Number14808233
Product Code MIH
UDI-Device Identifier00818009015105
UDI-Public(01)00818009015105(17)180511
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/11/2020
Device Model NumberBEA22-70/I20-30
Device Lot Number1685410-010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer01/26/2022
Initial Report FDA Received Date01/10/2022
Supplement Report FDA Received Date02/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/31/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA SUPRARENAL 1599296-025; AFX VELA SUPRARENAL 2308012-007; AFX2 BIFURCATED STENT GRAFT 1599020-007; OVATION IX EXTENDER FS110919-24; OVATION IX EXTENDER FS120419-08
Outcome Attributed to Adverse Event Other;
Patient Age80 YR
Patient SexMale
-
-