• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL LP; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL LP; HCG, KRD Back to Search Results
Model Number RBY4C0305-B
Medical Device Problem Code Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the gastric artery using ruby coil lps, an lp system detachment handle (handle) and a non-penumbra microcatheter.It was reported that the patient anatomy was tortuous.During the procedure, a ruby coil lp would not advance from its initial position within the introducer sheath, and therefore, the ruby coil lp was removed.Next, the physician advanced another ruby coil lp to the target location and attempted to detach it using the handle; however, the ruby coil lp failed to detach.The physician then tried detaching the coil manually, but the ruby coil lp still failed to detach.Subsequently, the physician decided to remove the microcatheter with the ruby coil lp inside.While retracting the microcatheter, the ruby coil lp unintentionally detached in the hepatic artery.A snare device was used to retrieve the coil; however, the coil could not be retrieved.The ruby coil lp was left in the hepatic artery.The procedure ended at this point.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL LP
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key13219471
Report Number3005168196-2022-00006
Device Sequence Number5365545
Product Code HCG
UDI-Device Identifier00814548018317
UDI-Public00814548018317
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K192955
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/10/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRBY4C0305-B
Device Catalogue NumberRBY4C0305
Device Lot NumberF109024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2021
Initial Report FDA Received Date01/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexFemale
Date Report Sent to FDA01/01/2005
-
-