| Model Number |
SONATA |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user reported that the hearing performance has decreased.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information from the field, the device does not provide sufficient benefit due to a partial migration of the active electrode out of cochlea, which could be confirmed by diagnostic imaging.A re-insertion surgery is planned in (b)(6) 2022.
|
| |
|
Event or Problem Description
|
|
The user reported that the hearing performance has decreased since the beginning of (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information from the field, the device does not provide sufficient benefit due to a partial migration of the active electrode out of cochlea, which could be confirmed by diagnostic imaging.Reportedly the recipient was re-fitted and is doing better.
|
| |
|
Event or Problem Description
|
|
The user reported that the hearing performance has decreased since the beginning on (b)(6) 2021.
|
| |
|
Event or Problem Description
|
|
The user reported that the hearing performance has decreased since the beginning of (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information from the field, the device did not provide sufficient benefit due to a partial migration of the active electrode out of cochlea, which could be confirmed by diagnostic imaging.A re-insertion surgery was performed successfully.The device remains implanted and in use.
|
| |
|