MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIMA BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIMA BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00539200
Medical Device Problem Codes Break (1069); Positioning Failure (1158); Material Deformation (2976); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter address 1: (b)(6).(b)(4).The returned flexima biliary stent was analyzed, and a visual evaluation noted that the stent attached to the push catheter by the suture thread.The push catheter had several kinks.The guide catheter had one part detached.During microscope inspection it was observed that the suture hole of the push catheter was torn.No other issues with the device were noted.The reported event was confirmed.It is most likely that the push catheter got kinked by user manipulation, some technique performed during procedure and/or even tortuous anatomy could have contributed on this issue, once the push catheter was kinked the user may have experienced difficulties to pull out the guide catheter and this may have taken the user to apply an extra force/tension, as consequence of the tough movement between the guide catheter and push catheter the suture hole got torn and the guide catheter got detached one part of it.Therefore, adverse event related to procedure is selected as the most probable root cause for the complaint.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation that a flexima biliary stent was used during an endoscopic retrograde cholangiopancreatography procedure in the bile duct, performed on (b)(6) 2021.During the procedure, the stent failed to deploy.Another flexima biliary stent was opened and successfully completed the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.This event has been deemed a reportable event based on the investigation finding of guide catheter break.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIMA BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13225207
Report Number3005099803-2021-08159
Device Sequence Number3787209
Product Code FGE
UDI-Device Identifier08714729337362
UDI-Public08714729337362
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K965147
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/28/2024
Device Model NumberM00539200
Device Catalogue Number3920
Device Lot Number0027054530
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/14/2021
Initial Report FDA Received Date01/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age45 YR
Patient SexFemale
Patient Weight65 KG
-
-