MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Overheating of Device (1437); Charging Problem (2892); Connection Problem (2900)
|
Patient Problem
Burn(s) (1757)
|
Event Date 12/29/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant products: product id: 97755, serial#: (b)(4), product type: recharger.Other relevant device(s) are: product id: 97755, serial/lot #: (b)(4), udi#: (b)(4).Date of event: date inaccurate, only the month and year are valid.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins).It was reported that the patient began having issues initiating a charge session to charge their ins.They had to manipulate the recharger cord where the cord met the paddle in order to initiate a charge session.The relay box of the recharger became so hot that it left a box shaped burn/red mark on their skin.The patient was recharging their ins and the controller battery depleted and their ins did not recharge at all.A replacement recharger was requested.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 97755 lot# serial# (b)(6) implanted: explanted: product type recharger medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the patient.The patient said the replacement recharger resolved the burning issue.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 97755 lot# serial# (b)(6) implanted: explanted: product type recharger.H3: product id 97755 serial# (b)(6) implanted: product type recharger.The recharger was returned for analysis.Analysis found that the recharger failed both plexus and bench testing and scrapped.The patient sent a replacement recharger.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|