| Model Number |
70805-70 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of 23 mg/dl and 236 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned as of this report.A valid meter serial number was not provided.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.A tripped trend review was conducted for the reported complaint and fs lite meter, no trends were identified that would indicate any product related issues.Dhrs for the freestyle test strips were reviewed and the dhrs showed the freestyle test strips passed all tests prior to release.Retain testing was performed for freestyle strips and all units performed in specification and passed.If the product is returned, the case will be re-opened and a physical investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|