• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. FASTTHREAD INTERFERENCE SCRW DRIVER, 6MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. FASTTHREAD INTERFERENCE SCRW DRIVER, 6MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number FASTTHREAD INTERFERENCE SCRW DRIVER, 6MM
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an anterior cruciate ligament surgery multiple screw drivers were blunt.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.26-oct-2021 update: further information were provided that the reported event occurred during an anterior cruciate ligament repair and the screwdriver was twisted and not dull.During returned device evaluation, a reportable malfunction was discovered.
 
Additional Manufacturer Narrative
During returned device evaluation, a reportable malfunction was discovered.Complaint confirmed, the tip was found to be damaged and broken.The tip id is also obstructed with unknown white material.The cause of the event is undetermined, however a likely cause is user-applied mechanical forces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FASTTHREAD INTERFERENCE SCRW DRIVER, 6MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13225808
Report Number1220246-2022-04275
Device Sequence Number17077225
Product Code LXH
UDI-Device Identifier00888867291928
UDI-Public00888867291928
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFASTTHREAD INTERFERENCE SCRW DRIVER, 6MM
Device Catalogue NumberAR-4019D
Device Lot Number47321824
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2021
Initial Report FDA Received Date01/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/20/2018
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-