| Model Number |
11.0 MM FLIP CUTTER II |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament (acl) all-inside surgery the device broke off in the patient, in the middle of backward drilling.The broken piece was retrieved from patient.No harm for patient, operator or third party was reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint is confirmed based on the customer provided photo, which displays the broken cutting tip.However, without return of the device for physical evaluation, the cause remains undetermined.No change in harm was identified.
|
| |
|