• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. TENO SCRW,BIO-COMP; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. TENO SCRW,BIO-COMP; SCREW, FIXATION, BONE Back to Search Results
Model Number TENO SCRW,BIO-COMP
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 12/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during an ucl reconstruction on the thumb the tenodesis screw broke off during insertion.The broken piece remains inside the patient.Metal rod is also installed in the patient.The surgery was finished successfully with the same device was used anyway.It was not necessary to switch the surgical technique or do a second surgery.Update 03-jan-2021: the implant broke after full insertion to intended depth.As the screw on the ar-1530bc is sitting over a small metal rot, when the item broke inside the patient, a piece of the metal rod also broke off and is remaining in the screw in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENO SCRW,BIO-COMP
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13225859
Report Number1220246-2022-04282
Device Sequence Number19766629
Product Code HWC
UDI-Device Identifier00888867018600
UDI-Public00888867018600
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K051726
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 01/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model NumberTENO SCRW,BIO-COMP
Device Catalogue NumberAR-1530BC
Device Lot Number13029071
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/23/2021
Initial Report FDA Received Date01/11/2022
Date Device Manufactured04/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-