ARTHREX, INC. TENO SCRW,BIO-COMP; SCREW, FIXATION, BONE Back to Search Results
Model Number TENO SCRW,BIO-COMP
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/22/2021
Type of Reportable Event Malfunction
Event or Problem Description
On 12/23/2021, it was reported by a arthrex employee via e-mail that a ar-1530bc bio-composite-tenodesis inserter broke.This occurred on (b)(6) 2021 during an unknow procedure after inserting the screw into the bone socket, the inserter was broke.No additional information provided, additional information requested.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 12/23/2021, it was reported by a arthrex employee via e-mail that a ar-1530bc bio-composite-tenodesis inserter broke.This occurred on (b)(6) 2021 during an unknow procedure after inserting the screw into the bone socket, the inserter was broke.No additional information provided, additional information requested.
 
Additional Manufacturer Narrative
Complaint confirmed.Visual inspection identified that the distal tip had broken off, with the drive geometry no longer present.Additionally, it was identified that approximately 23.38mm of the shaft had fragmented.Device testing could not be performed due to the damaged of the device.The cause of this event is undetermined; however, a probable cause can be attributed to prying/leveraging the driver tip within the screw hex during insertion.The method used to prepare the bone during the procedure, as well as the bone quality encountered, were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENO SCRW,BIO-COMP
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13225893
Report Number1220246-2022-04281
Device Sequence Number3734320
Product Code HWC
UDI-Device Identifier00888867018600
UDI-Public00888867018600
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K051726
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model NumberTENO SCRW,BIO-COMP
Device Catalogue NumberAR-1530BC
Device Lot Number13288354
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/23/2021
Supplement Date Received by Manufacturer12/23/2021
Initial Report FDA Received Date01/11/2022
Supplement Report FDA Received Date07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-