|
SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 5-6 10MM RT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
74035252 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference case-(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during field inspection, it was found that the jrny ii bcs art isrt trl sz 5-6 10mm rt is broken.No case involved; therefore, there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the associated device was returned and evaluated.The manufacturing process of the device did not contribute to the reported event.A visual inspection of the returned device confirmed the stated failure mode.The device is chipped rendering the device inoperative.A review of complaint history revealed 1 similar event for the listed part number.No adverse trend was found for this event.For the batch number, the review did reveal 2 similar events.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|