This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, the cause that led to connecting tube coating peeling off was surmised as follows: 1) improper reprocessing by customer.2) non-olympus approved chemicals used during reprocessing by customer.Specific probable causes may be the following: a.Physical stress: 1.) insertion section was contacted and rubbed by mouthpiece; 2.) the subject device was used for procedure while the insertion section was not smoothly passed through endotracheal tube; b.Chemical stress: 1.) water-insoluble anesthetic spray for pharyngeal area, such as xylocaine spray, was directly sprayed to the insertion section; 2.) the subject device was reprocessed with disinfectant which is not olympus recommendation.The instruction manual identifies the following related verbiage which may help prevent the phenomenon: ¿inspection of the endoscope : inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.¿ additionally, the user can reduce and reduce/prevent occurrence of the suggested event by handling device in accordance with the following ifu statement and item: ¿operation: insertion or withdrawal with excessive force, or forcible insertion or withdrawal; operation: insertion of the endoscope: if necessary, apply a medical-grade, water-soluble lubricant to the insertion tube; compatible reprocessing methods and chemical agents : olympus endoscopic equipment is compatible with several methods of reprocessing.Certain components and accessories, however, are not compatible with some methods, which can cause equipment damage.For appropriate reprocessing methods, refer to table 2.1 below, the recommendations of your infection control committee and all national and local hospital guidelines and policies.¿ olympus will continue to monitor field performance for this device.
|