Model Number 275010140
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/01/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2021 during pedicle probe step of procedure to ensure there isn¿t a pedicle breach, the end tip of the pedicle probe broke off in the patients l5 vertebral body.Procedure completed successfully without any surgical delay.This report is for one (1) ball tip feeler straight.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
This report is for one (1) ball tip feeler curved.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D4: lot number.H3, h6: a product investigation was conducted.The product was returned to depuy synthes cq for evaluation.The depuy synthes cq team conducted a visual inspection of the returned device.Visual analysis of the returned sample as well as the photo revealed that the ball tip feeler curved was broken in two places.The feeler shaft was broken off at the proximal connection point to the handle, and at the distal end near the ball.The distal end was not returned.The uploaded photo shows a broken off portion of the device embedded in the patient.No other issues were identified.The dimensional inspection was performed on the returned device and it was measured to be within specification the observed condition ball tip feeler curved in the device was consistent with a random component failure that may have been caused by exposure to unintended force.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the ball tip feeler curved was observed to be broken, and a broken off portion was shown embedded in the patient.While no definitive root cause could be determined, it is probable that the ball tip feeler curved broke due to unexpected forces.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Manufacturing record evaluation (mre) was not performed since no manufacturing related potential cause is suspected.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.B5; d1.D4: corrected catalog number and udi number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALL TIP FEELER CURVED
Common Device Name
PROBE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key13226818
Report Number1526439-2022-00071
Device Sequence Number4111710
Product Code HXB
UDI-Device Identifier10705034194066
UDI-Public(01)10705034194066
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number275010140
Device Catalogue Number275010160
Device Lot Number0705V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/15/2021
Supplement Date Received by Manufacturer02/03/2022
Initial Report FDA Received Date01/11/2022
Supplement Report FDA Received Date02/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-