• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2H8519
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient was a (b)(6).Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the tubing of a clearlink system non-dehp continu-flo solution set fell apart at the proximal 'y' site (clearlink component).This issue was identified during patient infusion.The set was connected to a large volume infusion pump running at a low rate.Blood backed up two inches in the tubing but did not reach the break in the tubing.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
Correction made to e1: initial reporter facility name: (b)(6).The device was received for evaluation.A visual inspection was performed and the tubing was observed separated from the second y- site clearlink in the upstream part.During further evaluation the solvent was observed not to be uniformly applied in the tubing.Dimensional test was performed, at the tubing was found to be within the product specification.The reported condition was verified.The cause of the condition was due to improper handling during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13226981
Report Number1416980-2022-00015
Device Sequence Number4101023
Product Code FPA
UDI-Device Identifier00085412048451
UDI-Public(01)00085412048451
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K961225
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number2H8519
Device Lot NumberR21F12031
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2021
Supplement Date Received by Manufacturer02/23/2022
Initial Report FDA Received Date01/11/2022
Supplement Report FDA Received Date02/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LARGE VOLUME INFUSION PUMP; UNSPECIFIED SOLUTION
Patient SexUnknown
-
-