| Model Number |
133S-MV-13-T |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified: opening, crease flat.Leak test was performed and found opening on the device.A microscopic analysis was performed which identify an opening striated in the radios.Based on the device analysis the final assessment is: a striated opening in the radios side assessed as surgical damage.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.The device has been explanted.
|
| |
|