Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Codes Electric Shock (2554); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) exhibited a gradual increase of high shock impedance measurements in the right ventricular (rv) lead over the past year.Data was sent to boston scientific technical services (ts) for their review and recommendations.Data was reviewed and troubleshooting was recommended by ts.Physician decided to schedule a new follow up.At this time, the system remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) exhibited a gradual increase of high shock impedance measurements in the right ventricular (rv) lead over the past year.Data was sent to boston scientific technical services (ts) for their review and recommendations.Data was reviewed and troubleshooting was recommended by ts.Physician decided to schedule a new follow up.At this time, the system remains in service.No adverse patient effects were reported.Additional information received confirmed that physician performed induction test.Impedance measurement were (b)(4).No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE SG
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13228488
Report Number2124215-2021-39364
Device Sequence Number14494342
Product Code NVY
UDI-Device Identifier00802526431388
UDI-Public00802526431388
Combination Product (Y/N)Y
Initial Reporter CountryIT
PMA/510(K) Number
P910073/S077
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/27/2013
Device Model Number0283
Device Catalogue Number0283
Device Lot Number100771
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/09/2021
Supplement Date Received by Manufacturer01/26/2022
Initial Report FDA Received Date01/11/2022
Supplement Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/27/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-