| Model Number |
82-6631 |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/12/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
A facility reported a microsensor was implanted on (b)(6) 2021, three days later it could not be detected and was explanted on (b)(6) 2021.Patient is in stable condition.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The microsensor (b)(4) was returned for evaluation.Failure analysis - received catheter in a drainage tube.Debris covering sensor area and suture right below sensor.Sensor diaphragm is broken.Icp express read no transducer detected.No testing was possible.The issue of the complaint has been confirmed.The possible root cause of the defective device could be due to mishandling of the catheter.
|
| |
|