Model Number 82-6631
Medical Device Problem Code No Device Output (1435)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/12/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported a microsensor was implanted on (b)(6) 2021, three days later it could not be detected and was explanted on (b)(6) 2021.Patient is in stable condition.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The microsensor (b)(4) was returned for evaluation.Failure analysis - received catheter in a drainage tube.Debris covering sensor area and suture right below sensor.Sensor diaphragm is broken.Icp express read no transducer detected.No testing was possible.The issue of the complaint has been confirmed.The possible root cause of the defective device could be due to mishandling of the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROMONITOR BASIC KIT
Common Device Name
ICP MICROSENSORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13228993
Report Number3013886523-2021-00594
Device Sequence Number19874881
Product Code GWM
UDI-Device Identifier20886704042317
UDI-Public20886704042317
Combination Product (Y/N)N
PMA/510(K) Number
K914479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number82-6631
Device Catalogue Number826631
Device Lot Number3963073
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/26/2021
Supplement Date Received by Manufacturer03/15/2022
Initial Report FDA Received Date01/11/2022
Supplement Report FDA Received Date03/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1