WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.0MM TI LOCKING SCREW W/T25 STARDRIVE 44MM FOR IM NAILS; SCREW, FIXATION, BONE Back to Search Results
Model Number 04.005.434
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 01/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: additional patient identifier: (b)(6).Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, the patient underwent a hardware removal due to nonunion at fracture site.Surgeon removed an ex tibial nail and 4 interlocking screws.Patient was revised to another companies products.No further information provided.This report is for one (1) 4.0mm ti locking screw w/t25 stardrive 44mm for im nails this is report 3 of 5 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6- the complaint device was not received for investigation.The images were reviewed, and the complaint condition is not confirmed.After review of the x-rays provided, one tibial nail is present with no signs of issue with it, it is located in the tibial area.The nail does not appear to have any defect or malfunction that could contribute to the nonunion condition.This complaint is not confirmed as a photo inspection does not show malfunction of the complaint device.A manufacturing record evaluation cannot be performed due to lack of lot number.A definite assignable root cause could not be determined based on the provided information.As the device was not returned, and as-received condition could not be assessed and a dimensional inspection and document/specification review were not completed.No dhr review was completed since no lot number was supplied via photo or additional information.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot = the device lot number is unknown, therefore a device history review could not be performed.If more information become available, the record will be re-assessed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.0MM TI LOCKING SCREW W/T25 STARDRIVE 44MM FOR IM NAILS
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key13232085
Report Number2939274-2022-00125
Device Sequence Number4547662
Product Code HWC
UDI-Device Identifier10886982085467
UDI-Public(01)10886982085467
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K103002
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.005.434
Device Catalogue Number04.005.434
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/16/2021
Supplement Date Received by Manufacturer02/03/2022
Initial Report FDA Received Date01/11/2022
Supplement Report FDA Received Date02/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4.0 TI LCKNG SCR T25 SD 44 FOR IM NAILS; 4.0 TI LCKNG SCR T25 SD 44 FOR IM NAILS; 4.0 TI LCKNG SCR T25 SD 44 FOR IM NAILS; 9 TI CANN TIBIAL NAIL-EX PROX BEND 375-S
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient SexMale
-
-