• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL TECHNOLOGY, INC. MINI ONE(R) BALLOON BUTTON; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL TECHNOLOGY, INC. MINI ONE(R) BALLOON BUTTON; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number M1-5-1410-I
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is a response to uf report # (b)(4), which was received by amt from the fda on 12/07/2021.Based on the provided information, the complaint is not a reportable event per 21 cfr section 803.There was no death.There was no serious injury, as defined by the fda.This was not life-threatening, nor did this result in permanent impairment or necessitate medical or surgical intervention to preclude permanent impairment.Amt confirmed that the device referenced in the report is not available for return.Since the device was not returned, a visual and functional evaluation could not be performed and device failure cannot be confirmed at this time.A device history review was completed for the reported lot number and no anomalies were found and there have been no other complaint reports from this same batch.The complaint information has been logged into our complaint database for trending purposes.Complaint # (b)(4) was assigned to this report.
 
Event or Problem Description
Based on the original reporter's provided report # (b)(4), "the internal retention balloon of the 14fr x 1.0cm minione® balloon button low profile feeding device leaked causing the gastrostomy tube to be dislodged from the patient.Patient's mother reported this as the second time this has happened, a similar situation occurred two weeks prior where the tube came out and the balloon was not intact.When trying to inflate the tube there was obvious leakage noted (lot 210715-116, expiration date 2024-06-01, ref # m1-5-1410-i).".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI ONE(R) BALLOON BUTTON
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
APPLIED MEDICAL TECHNOLOGY, INC.
8006 katherine boulevard
brecksville OH 44141
Manufacturer (Section G)
APPLIED MEDICAL TECHNOLOGY, INC.
8006 katherine boulevard
brecksville OH 44141
Manufacturer Contact
joshua meinke
8006 katherine boulevard
brecksville, OH 44141
4407174000
MDR Report Key13232815
Report Number1526012-2021-00021
Device Sequence Number17218257
Product Code KNT
UDI-Device Identifier00842071113940
UDI-Public(01)00842071113940(17)240601(10)210715-116
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K161413
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date06/01/2024
Device Model NumberM1-5-1410-I
Device Catalogue NumberM1-5-1410-I
Device Lot Number210715-116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 12/07/2021
Initial Report FDA Received Date01/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age5 MO
Patient SexMale
-
-