| Model Number |
SSFT215PRE |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unruptured aneurysm endovascular coil embolization procedure, after flushing the introducer sheath, when the surgeon inserted and pushed forward the subject guidewire a little resistance was felt in the middle.When the introducer sheath was advanced to the end of the guidewire, the tip of the introducer sheath was covered with a green fibrous foreign body.So, the surgeon replaced the subject guidewire with a new device.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
There are controls in the manufacturing process to ensure the product met specifications upon release.Visual/microscopic inspection: an attempt had been made to shape the guidewire tip.The guidewire ptfe coating was noted to be peeling to 7cm from the proximal end.The guidewire was returned with an introducer.Ptfe coating peelings were noted in both ends of the introducer.The core wire distal end was observed through the distal tip dome.Hydrophilic coating was hydrated there were no anomalies noted.Functional test: the guidewire was advanced through a flushed demonstration sl - 10 microcatheter, no anomalies were noted.The event was confirmed based on the damage noted to the returned device.The device failed to meet specification when returned for analysis.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Additional information indicates the issue was noted while transferring the guidewire into the hub of the microcatheter using introducer sheath while preparation outside the patient.Analysis of the returned device found an attempt had been made to shape the guidewire tip.The guidewire was returned with the introducer.The guidewire ptfe coating was noted to be peeling to 7cm from the proximal end.The introducer was noted to have evidence of ptfe peeling on either ends.The peeling most likely occurred as a result of interaction with another device.The damage noted is consistent with backloading the proximal end of the guidewire into the distal end of the introducer and pulling it through the introducer.The as reported and as analyzed ¿guidewire ptfe coating peeling¿ will be assigned handling damage as the issue is due to handling of the product or portion of the product during the clinical procedure, upon removal of the product from the packaging, or preparation of the product prior to use.The reported defect guidewire friction was not confirmed during the analysis.It is possible that the ptfe scrapings noted may have contributed to the reported defect but this was not noted during the analysis of the returned device.An assignable cause of 'not confirmed' will be assigned as the issue included a returned product review which showed no evidence of either the alleged issue or any defect which could have contributed to the event.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unruptured aneurysm endovascular coil embolization procedure, after flushing the introducer sheath, when the surgeon inserted and pushed forward the subject guidewire a little resistance was felt in the middle.When the introducer sheath was advanced to the end of the guidewire, the tip of the introducer sheath was covered with a green fibrous foreign body.So, the surgeon replaced the subject guidewire with a new device.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
|
| |
|
Search Alerts/Recalls
|