Event or Problem Description
It was reported that an intraocular lens (iol) was explanted from the patients left eye due to patient experiencing glare and a thumbprint constantly over her vision.The iol was replaced with a non-johnson & johnson lens.There were no surgical or medical interventions such as incision enlargement, capsule tear, vitrectomy or sutures.There was no patient injury.Reportedly, the patient is doing ok post-operatively.No further information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis, therefore a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation? yes.Section d9: date returned to manufacturer: jan 12, 2022.Section h3: evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed viscoelastic residue on the optic body and haptics and that the lens was received cut in half, which is consistent with a lens that was handled during explant.Based on the return condition of the lens, no further evaluation could be performed.The complaint issue could not be confirmed a product quality deficiency could not be determined.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no other complaints were received for this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13233657
Report Number3012236936-2022-00096
Device Sequence Number4727236
Product Code MFK
UDI-Device Identifier05050474751187
UDI-Public(01)05050474751187(17)240307
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/07/2024
Device Model NumberDFW150
Device Catalogue NumberDFW150U160
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2021
Supplement Date Received by Manufacturer02/11/2022
Initial Report FDA Received Date01/11/2022
Supplement Report FDA Received Date02/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight95 KG
Patient RaceWhite
-
-