• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRIO ST; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRIO ST; SURGICAL MESH Back to Search Results
Catalog Number 5950020
Medical Device Problem Codes Defective Device (2588); Patient-Device Incompatibility (2682); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Adhesion(s) (1695); Scar Tissue (2060); Hernia (2240); Obstruction/Occlusion (2422); Insufficient Information (4580)
Date of Event 01/30/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.This emdr represents the bard/davol ventrio st (device #2).An additional emdr was submitted to represent the bard/davol ventralex st (device #1).Should additional information be provided, a supplemental emdr will be submitted.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of unspecified bard/davol ventralex st patch and ventrio st on (b)(6) 2013.As reported, the patient is making a claim for an adverse patient outcome against both devices.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of unspecified bard/davol ventralex st patch and ventrio st on (b)(6) 2013.As reported, the patient is making a claim for an adverse patient outcome against both devices.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6) 2013 ¿ patient was diagnosed with incisional hernia and umbilical hernia thereby underwent open repair with the implant of ventralex st mesh (device 1) and ventrio st mesh (device 2).Per operative notes, ¿a ventralex st mesh (device 1) was placed into umbilicus and a ventrio st mesh (device 2) was placed to right upper quadrant incisional hernia using sutures.¿ (b)(6) 2020 ¿ patient was diagnosed with incarcerated spigelian ventral hernia; bowel obstruction thereby underwent open repair with the removal of mesh (device 2).Per operative notes, ¿prior incisional hernia scar was noted.Adhesions were taken down.The old mesh (device 2) stuck to the bowel was removed.¿.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Addendum: this supplemental emdr is submitted to document additional information provided.Root cause is inconclusive, no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-mar-2013) is considered to be a best estimate.This supplemental emdr represents the bard/davol ventrio st (device 2).An additional emdr was submitted to represent the bard/davol ventralex st (device 1).Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRIO ST
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key13235217
Report Number1213643-2022-90162
Device Sequence Number4598976
Product Code FTL
UDI-Device Identifier00801741031519
UDI-Public(01)00801741031519(17)150328(10)HUXC1014
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K101920
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/28/2015
Device Catalogue Number5950020
Device Lot NumberHUXC1014
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/10/2022
Supplement Date Received by Manufacturer08/21/2025
Initial Report FDA Received Date01/12/2022
Supplement Report FDA Received Date09/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/15/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient SexFemale
-
-