Model Number 1365-07-000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Inflammation (1932); Pain (1994); Pocket Erosion (2013); Distress (2329); Metal Related Pathology (4530); Unspecified Tissue Injury (4559); Physical Asymmetry (4573); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 08/24/2020
Type of Reportable Event Serious Injury
Event or Problem Description
Litigation alleges toxic levels of metal ions in the tissue, bone, joint, blood stream causing destruction, pain and emotional distress.Doi: (b)(6) 2008.Dor: (b)(6) 2020.Left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Initial reporter occupation: lawyer.(b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.H6 clinical code: unspecified tissue injury (e2015) is used to capture soft tissue injury and bone injury.
 
Event or Problem Description
After review of the operative notes and sticker sheets, patient was revised due to metallosis, pain similar trochanteric bursitis.The patient had a leg length discrepancy to begin with.Patient stated he felt like it was about a quarter of an inch when he was asleep.The hip was stable and has erosion at the head and neck junction.Doi: (b)(6) 2010, dor: (b)(6) 2020, affected side: left hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12/14 ARTICUL 40MM M SPEC+8.5
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13235259
Report Number1818910-2022-00576
Device Sequence Number16783337
Product Code LPH
UDI-Device Identifier10603295033004
UDI-Public10603295033004
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K060031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/11/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2014
Device Model Number1365-07-000
Device Catalogue Number136507000
Device Lot Number2880671
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2021
Supplement Date Received by Manufacturer03/08/2023
Initial Report FDA Received Date01/12/2022
Supplement Report FDA Received Date03/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/06/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INNACLE MTL INS NEUT40IDX56OD; PINN CAN BONE SCREW 6.5MMX30MM; PINNACLE SECTOR II CUP 56MM; TRI-LOCK BPS SZ 5 HI OFFSET; UNK HIP ACETABULAR LINER METAL PINNACLE; UNKNOWN HIP FEMORAL STEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexMale
-
-