STRYKER GMBH EXTRACTOR ASNIS III FOR SCREWS:Ø4.0MM AO FITTING; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 702496
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
As reported: "during removal, the screw stopped in the middle and the tip of the extractor broke off when the surgeon kept turning it.The bone around the screw was then drilled using a crown drill and the screw was removed.In doing so, the drill tip may have hit the screw and chipped it.".
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the tip of the extractor is broken off as complained.The device inspection revealed the following: the visual inspection has shown that the tip of the extractor is broken off, the fragment was not returned for evaluation.The fracture face runs diagonal, which indicates high lateral forces during the breakage.The fracture face has the view of a ductile overload fracture, where first a permanent distortion of the material occurred which did finally end in the breakage of the tip.Furthermore, a hardness test according to rockwell on the returned item indicates the material being in accordance to the values in the material specification.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to be user related.The breakage shows typical signs of torsional and bending overload by excessive forces applied on the extractor during the extraction of the screw.If more information is provided, the case will be reassessed.
 
Event or Problem Description
As reported: "during removal, the screw stopped in the middle and the tip of the extractor broke off when the surgeon kept turning it.The bone around the screw was then drilled using a crown drill and the screw was removed.In doing so, the drill tip may have hit the screw and chipped it.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRACTOR ASNIS III FOR SCREWS:Ø4.0MM AO FITTING
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
sharon rivas
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13235386
Report Number0008031020-2022-00008
Device Sequence Number11839175
Product Code LXH
UDI-Device Identifier04546540112415
UDI-Public04546540112415
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number702496
Device Catalogue Number702496
Device Lot Number00257V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/15/2021
Supplement Date Received by Manufacturer01/27/2022
Initial Report FDA Received Date01/12/2022
Supplement Report FDA Received Date02/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-