Model Number 7311
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
 
Event or Problem Description
The customer reported that she obtained blood glucose readings of 516 mg/dl at 11:22 a.M., 65 mg/dl at 11:23 a.M., 180 mg/dl at 11:27 a.M., and 105 mg/dl at 11:29 a.M.With the contour next link 2.4 meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.Replacement meter kit and test strips were sent to the customer.
 
Additional Manufacturer Narrative
Based on the clarification received, the customer's date of birth and age were updated in section a2.
 
Additional Manufacturer Narrative
The customer did not return the suspected device for evaluation.Therefore, the in-house contour next link 2.4 meter was tested with the in-house contour next test strips from lot # 1bpeg18a and in-house contour next control solution.Additionally, the in-house meter was also tested with the in-house test strips using blood sample.All results obtained during in-house testing were within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT
Common Device Name
BLOOD GLUCOSE TEST STRIPS
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
Manufacturer (Section G)
PHC CORPORATION (REGISTRATION NUMBER: 1000134386)
2131-1 minamigata
toon ehime, 791-0 395
JA   791-0395
Manufacturer Contact
shweta gulati
100 summit lake drive
valhalla, NY 10595
9142361830
MDR Report Key13235844
Report Number1810909-2022-00005
Device Sequence Number4727812
Product Code NBW
UDI-Device Identifier10301937311502
UDI-Public10301937311502
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P150001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/28/2023
Device Model Number7311
Device Catalogue Number7311
Device Lot Number1BPEG18A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/29/2021
Supplement Date Received by Manufacturer01/24/2022
05/12/2022
Initial Report FDA Received Date01/12/2022
Supplement Report FDA Received Date02/02/2022
05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age55 YR
Patient SexFemale
-
-