Model Number 1806-4265
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
It was reported: "the drill shaft broke when it was used.The broken tip was removed from the patient during the surgery.".
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned for evaluation, and matches the alleged failure.The device inspection revealed the following: the received drill shows signs of intense usage which is evident from the cutting edges which are blunt and slightly damaged.The drill was found to be broken from around the middle of the flute region.The breakage surface reveals a smooth surface.This confirms a sudden brittle fracture which was caused due to the application of high bending and torsional load.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of manufacturing or design-related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the investigation, the root cause was attributed to being user related.The appearance of breakage surface and cutting edges indicates an intense usage and application of high bending and torsional load, leading to a forced brittle fracture.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported: "the drill shaft broke when it was used.The broken tip was removed from the patient during the surgery.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL, TRI-FLAT T2 FEMUR Ø4,2X340 MM
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
marilyne chaumont
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13236093
Report Number0009610622-2022-00012
Device Sequence Number4598999
Product Code HWE
UDI-Device Identifier04546540196781
UDI-Public04546540196781
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1806-4265
Device Catalogue Number18064265
Device Lot NumberK106036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/16/2021
Supplement Date Received by Manufacturer03/17/2022
Initial Report FDA Received Date01/12/2022
Supplement Report FDA Received Date04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-