| Model Number |
1806-4265 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of the investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported: "the drill shaft broke when it was used.The broken tip was removed from the patient during the surgery.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, since the device was returned for evaluation, and matches the alleged failure.The device inspection revealed the following: the received drill shows signs of intense usage which is evident from the cutting edges which are blunt and slightly damaged.The drill was found to be broken from around the middle of the flute region.The breakage surface reveals a smooth surface.This confirms a sudden brittle fracture which was caused due to the application of high bending and torsional load.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of manufacturing or design-related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the investigation, the root cause was attributed to being user related.The appearance of breakage surface and cutting edges indicates an intense usage and application of high bending and torsional load, leading to a forced brittle fracture.If any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported: "the drill shaft broke when it was used.The broken tip was removed from the patient during the surgery.".
|
| |
|
Search Alerts/Recalls
|