Event or Problem Description
|
|
Physician attempted to use a 3q microvascular plug during procedure.The mvp plug was soaked in heparinized saline prior to use.The catheter was flushed before inserting the plug.A 0.027" microcatheter was used for delivery.The target vessel diameter was 1.5-2.0 mm.There was no resistance felt while advancing the plug through the introducer sheath or catheter.The delivery wire was not manipulated or rotated before or during the delivery of the plug through the catheter.There was no kink or deformation noted on the device.There was no strut fracture observed.The user did not rotate the torquer in the incorrect direction accidentally during prep.It is reported the mvp plug was dislodged inside microcatheter.Physician removed the plug wire to position it better but realized that the mvp plug was dislodged inside microcatheter.Hence he removed the whole microcatheter out from patient and retrieved the plug out using a wire.The device was safely removed from the patient.The procedure was completed with coils.No injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: one image was provided for evaluation.The image is of the device inside a bio hazard bag.The micro catheter is loaded in the yellow delivery sheath.The mvp is loose in the bag and detached from the wire.The mvp plug is dislodged from the catheter but it is unclear from the image provided when the mvp dislodged from the catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|