| Model Number |
115-222 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: "rupture".
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side "rupture." device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: opening in the shell, underweight, crease flat and red particle in the shell.A microscopic analysis was performed which identify a sharp edge opening.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a opening sharp in the posterior side assessed as unidentified (tear) opening.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side rupture.Device was explanted.
|
| |
|