Model Number N/A
Medical Device Problem Codes Migration or Expulsion of Device (1395); Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Implant migration - blunt screw backed out: it was reported that after 2 weeks from the implantation, surgeon found #1 screw of the proximal screws was backed out from the proper position.
 
Additional Manufacturer Narrative
The manufacturer did not receive x-rays, or other source documents for review.The manufacturer did not receive the device for investigation.The lot number of the device was received.The device history records will be reviewed during investigation.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available an updated report will be submitted.Zimmer biomet¿s reference number of this file is (b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2022-00020, 0009613350-2022-00022, 0009613350-2022-00023.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Event description: it was reported that 2 weeks after the initial surgery, the surgeon found screw #1 of the proximal screws was backed out from the proper position.No revision is currently planned.Review of received data: no medical data relevant to the case has been received.Due diligence: no further due diligence required as all required information to support the conclusion is available or was already requested.Product evaluation: no product was returned; therefore, visual and dimensional evaluation could not be performed.Review of product documentation: device purpose: all involved devices are intended for treatment.Product compatibility: the product combination was approved by zimmer biomet.Dhr review: review of the device history records identified no deviations or anomalies during manufacturing.Review of the device history records for part # 47249616109 lot # 3058647 identified the two deviations and/or anomalies.Ncr(s): no ncr with a potential correlation to the reported event was found.Surgical technique identified that the interlocking screws should be locked using the corelock driver after the placement of all screws.It was noted that the surgeon followed the surgical technique.Conclusion: it was reported that 2 weeks after the initial surgery, the surgeon found screw #1 of the proximal screws was backed out from the proper position.No revision is currently planned.Based on the investigation the reported event cannot be confirmed.The investigation did identify a nonconformance or a complaint out of box (coob).Based on the given information and the results of the investigation, we were not able to identify a specific root cause for this issue.The need for corrective measures is not indicated and zimmer switzerland manufacturing gmbh considers this case as closed.Zimmer biomet¿s reference number of this file is (b)(4).
 
Event or Problem Description
No change to previously reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLUNT TIP SCREW, 4X34MM
Common Device Name
AFFIXUS(R) NATURAL NAIL(R) SYSTEM HUMERAL NAIL
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13237006
Report Number0009613350-2022-00021
Device Sequence Number17325375
Product Code HSB
UDI-Device Identifier00889024505414
UDI-Public00889024505414
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200814
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2025
Device Model NumberN/A
Device Catalogue Number47-2486-034-40
Device Lot Number3024655
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2021
Supplement Date Received by Manufacturer01/27/2022
Initial Report FDA Received Date01/12/2022
Supplement Report FDA Received Date02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-