MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX ST 100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX ST 100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 107636
Device Problems Disconnection (1171); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2021
Event Type  malfunction  
Event Description
It was reported that after priming one unit of a prismaflex st100 set, the post filter tubing disconnected from the filter and a saline leak was observed.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Initial reporter address:(b)(6).Prismaflex st100 set has been temporarily approved for use in the us under emergency use authorization (b)(4) to deliver crrt to treat patients in an acute care environment during the covid-19 pandemic.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
The actual device was provided for evaluation.The visual inspection showed that the return line was disconnected from the filter.The reported condition was verified.The cause could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX ST 100
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13237119
MDR Text Key283683365
Report Number8010182-2021-00450
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414075613
UDI-Public(01)07332414075613
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Catalogue Number107636
Device Lot Number20H1202
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/20/2021
Initial Date FDA Received01/12/2022
Supplement Dates Manufacturer Received02/09/2022
Supplement Dates FDA Received02/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-