This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.The reported failure mode will be monitored for future reoccurrence.Alleged failure: blurry image.Probable root cause: cables, connectors, digital board, power supply, filter / fuse, ac inlet board, main board, transition board, fiber board, intermittence between transition board and ch connector, fan / insufficient cooling, software, camera head (sensor, sensor cable), coupler (dust on optics, scratched optics, broken / shifted lenses, focusing mechanism, endoclamp) extension cable, intermittence while using rf devices (ex: valleylab), problem toggling light source from camera head, connected monitors ,leaking, manufacturing/ service nonconformity, use error.Manufacture date is not known h3 other text : 81.
|