• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC SOUNDSTAR® ECO DIAGNOSTIC ULTRASOUND CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC SOUNDSTAR® ECO DIAGNOSTIC ULTRASOUND CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number 10438577
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Ventricular Fibrillation (2130)
Date of Event 12/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Event or Problem Description
It was reported that a male patient born on (b)(6), ((b)(6)kg) underwent an atrial fibrillation (afib) ablation procedure with a soundstar® eco diagnostic ultrasound catheter.The patient suffered a vascular perforation.During an afib case, a perforation of the left femoral was discovered by flouro and contrast.The patient's blood pressure was stable and was under observation for about 20 minutes and no medical intervention in the current moment was provided.The patient was reported to be in stable condition.The physician believes that when advancing/enforcing of the intra-cardiac echocardiography (ice) catheter might have caused the perforation.The decanav was used after the ice catheter.It was also reported one of the current cable got stuck in a port on the been and when the patient got transferred the green patch cable got torn.The adverse event was discovered during the use of biosense webster products.The physician pushed too hard in the femoral vein with the ultrasound catheter without understanding the venous anatomy and perforated.He does not think it was a product malfunction, simply just forced a catheter without checking.The patient outcome of the adverse event is fully recovered.The patient did not require extended hospitalization because of the adverse event.Generator information: stockert smartablate, ref: m490002, sn: (b)(4).Since the event is life threatening and might result in permanent impairment of a body function or permanent damage of a body structure, it is to be considered serious and mdr-reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOUNDSTAR® ECO DIAGNOSTIC ULTRASOUND CATHETER
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
MDR Report Key13237965
Report Number3023245-2022-00001
Device Sequence Number19894525
Product Code OBJ
UDI-Device Identifier10846835008838
UDI-Public10846835008838
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/12/2022
Report Date (Section F) 01/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number10438577
Device Catalogue Number10438577
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/07/2022
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date01/12/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTO 3 SYSTEM.; DECANAV CATHETER.; SMARTABLATE GENERATOR.
Outcome Attributed to Adverse Event Life Threatening;
Patient Age71 YR
Patient SexMale
Patient Weight129 KG
Date Report Sent to FDA01/12/2022
-
-