Model Number 478001000
Health Effect - Clinical Codes Airway Obstruction (1699); Foreign Body In Patient (2687)
Date of Event 12/07/2021
Type of Reportable Event Malfunction
Event or Problem Description
An ascope 4 broncho was used in combination with a shiley evac oral tracheal tube.When the ascope 4 broncho large was pulled out, the tip of the bronchoscope teared off.The bending piece got stuck in the tube and caused an airway obstruction.The user has confirmed that the tracheal tube was bent during the procedure.
 
Additional Manufacturer Narrative
The tip of the bronchoscope tearing off was due to the user bending the et tube, causing this section of the ascope to not be in a straight position during removal of the scope.Thus, user applied excessive force to remove the scope from the tube causing the ascope bending section to tear off during removal of the scope.Further, according to the ifu for ascope 4 broncho family the user should always make sure that the bending section is in a straight position (neutral/non-deflected) and excessive force should not be used when withdrawing the endoscope.The product risk evaluation has been evaluated and the risk is assessed as acceptable.No remedial, corrective or preventive actions are taken as a result of this event.Ambu will keep monitoring this issue and take actions if necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMBU ASCOPE 4 BRONCHO LARGE
Common Device Name
FLEXIBLE VIDEO BRONCHOSCOPE
Manufacturer (Section D)
AMBUA/S
baltorpbakken 13
ballerup, dnk 2750
DA  2750
MDR Report Key13238803
Report Number1220828-2022-00001
Device Sequence Number20134151
Product Code EOQ
UDI-Device Identifier05707480135530
UDI-Public05707480135530
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number478001000
Device Catalogue Number478001000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/15/2021
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date01/12/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-