Model Number 3507555MC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Crushing Injury (1797)
Date of Event 09/13/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a (b)(6) hispanic female patient, who's undergone primary breast augmentation with two 555cc mentor memorygel breast implants, suffered right breast implant rupture, post-procedure.The patient was reportedly involved in an auto accident.The rupture was diagnosed via cat scan.As a result, the patient underwent bilateral removal and replacement on (b)(6) 2021.The replacements were: (right) 650cc mentor memorygel breast implant catalog: 3506504bc lot: 7666803 sn: (b)(4) and (left) 650cc mentor memorygel breast implant catalog: 3506504bc lot: 9526990 sn: (b)(4).
 
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: material rupture.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) hispanic female patient, who's undergone primary breast augmentation with two 555cc mentor memorygel breast implants, suffered right breast implant rupture, post-procedure.The patient was reportedly involved in an auto accident.The rupture was diagnosed via cat scan.As a result, the patient underwent bilateral removal and replacement on (b)(6) 2021.The replacements were: (right) 650cc mentor memorygel breast implant catalog: 3506504bc lot: 7666803 sn: (b)(4) and (left) 650cc mentor memorygel breast implant catalog: 3506504bc lot: 9526990 sn: (b)(4).
 
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: material rupture.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On (b)(6) 2022, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted the visual inspection.Visual analysis of the returned sample revealed that the sm mod classic gel, 555cc breast implant was found to be ruptured.The evaluation determined that the cause of the rupture is the trauma experienced.Rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Trauma is a known inherent risk of the gel-filled mammary prosthesis.It should be noted that as part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
gabriel alfageme
3041 skyway circle north
irving, TX 75038
9497898687
MDR Report Key13239328
Report Number1645337-2022-00411
Device Sequence Number4547879
Product Code FTR
UDI-Device Identifier00081317004257
UDI-Public00081317004257
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2018
Device Model Number3507555MC
Device Catalogue Number3507555MC
Device Lot Number6777810
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/20/2021
Supplement Date Received by Manufacturer01/31/2022
Initial Report FDA Received Date01/12/2022
Supplement Report FDA Received Date02/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/12/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient SexFemale
Patient EthnicityHispanic
-
-