The axium prime coil and was returned for analysis.The echelon-10 micro catheter used during the event was not returned.The axium prime coil was returned without the introducer sheath.The axium implant coil pushwire was found to be kinked at 32.5cm from proximal end.The axium implant coil was found to be broken.Based on the device analysis and reported information, the customer¿s report of ¿coil stretch¿ was unable to be confirmed as the coil was found to be broken.There was no reported issue pushing the axium prime coil out from within the introducer sheath.Therefore, it is possible the axium coil became damaged upon removal from within the introducer sheath; however, the root cause could not be determined.If information is provided in the future, a supplemental report will be issued.
|