| Model Number |
3549213 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
10/19/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient underwent a breast reconstruction revision with a 450cc mentor cpx 4 breast tissue expander with suture tabs and experienced a deflation on the left side post-operatively.As a result, the patient underwent explantation on (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient underwent a breast reconstruction revision with a 450cc mentor cpx 4 breast tissue expander with suture tabs and experienced a deflation on the left side post-operatively.As a result, the patient underwent explantation on (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2022, the mentor failure analysis lab received the device for evaluation.On (b)(6) 2022, mentor completed evaluation of the device.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection, leak testing, and microscopic examination of the returned device.Visual analysis of the returned sample revealed that the med height te w/sut tabs 450cc tissue expander was found to have a tear on the anterior view between the union of the shell and patch.Leak testing was performed, in accordance with mentor procedures, and no more leak sites were detected during the analysis.A microscopic examination was performed and the cause of the rupture could not be identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of tissue expanders include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: pc-(b)(4).
|
| |
|